Unholtz Dickie Corp

Model 8B6 Piezoelectric Accelerometer

The Model 8B6 accelerometer features a flat charge vs. frequency response as well as minimum charge sensitivity variation over a wide range of temperatures. The accelerometer design is a single-ended compression type with a double base construction to minimize base strain sensitivity. The Bismuth Titanate (PZ46) ceramic provides excellent sensitivity stability as well as flat temperature response.

Download Instrumentation Bulletins

Dynamic Performance

  • Charge sensitivity: 8.6 pC/g, ±30% (0.88 pC/m/s2)
  • Accelerometer capacitance: 1,500 pF typical at 160 Hz.
  • Mounted resonant frequency: 30,000 Hz. typical
  • Transverse sensitivity: 3% max at 160 Hz. in any direction, 1% typical
  • Accelerometer resistance: 20 megaohms min at 75º F, 500 megaohms min at 500º F
  • Frequency Max Deviation from Reference: 2Hz: 5%, 10Hz: 2%, 160Hz: REF, 1kHz: 2%, 5kHz: 5%, 7kHz: 7%, 10kHz: 15%
  • Amplitude linearity: within 2% of best straight line (2000 g range)

Specifications

Mechanical

Environmental

Calibration of each accelerometer is accomplished by comparison to a NIST traceable reference. Using a UD Model APEX SL-C Digital Calibration Control System, a continuous calibration curve is plotted from 10-10,000 Hz., showing percent deviation of charge sensitivity as compared to the sensitivity measured at the 160 Hz. reference frequency.